Blood Gas System
GASTAT-700Mode

Measured

Reagents Cal1,Cal2,Gas1,Gas2 one packages Flash works alone (Low Cost)

Test

No. of test unlimited

pH

6.000-8.000

PCO2, Torr

10.0-200.0

PO2, Torr

5.0-760.0

cHb, g/dL

tHb, g/dL

2.0-23.0

Ca++, mmol/L

0.5-5.0

Cl-, mmol/L

50.0-200.0

K+, mmol/L

1.00-20.00

Na+, mmol/L

50.0-200.0

Glucose, mg/dL

20.0-500.0

Lactate, mg/dL

0.4-30.0

tBilirubin, mg/dL

0.0-60.0

sO2, %

0.0-100.0

FO2Hb, %

0.0-100.0

FCOHb, %

0.0-100.0

FMetHb, %

0.0-100.0

FHHb, %

0.0-100.0

 

 

Calculated

 

PH(T)

Yes

pCO2(T)

Yes

pO2(T)

Yes

cHCO3-act

Yes

cHCO3-std

Yes

cBE(ecf)

Yes

cBE(B)

Yes

cBB

Yes

ctCO2(P)

Yes

ctCO2(B)

Yes

sO2(est)

Measured

BO2

Yes

p50

Yes

p50(T)

Yes

p50(st)

Yes

pO2/FiO2

Yes

pO2(A)

Yes

pO2(A) (T)

Yes

pO2(A-a)

Yes

pO2(A-a) (T)

Yes

pO2(a/A)

Yes

pO2(a/A) (T)

Yes

RI

Yes

RI(T)

Yes

ctO2(a)

Yes

ctO2(B)

Yes

ctO2(v)

Yes

ctO2(a-v)

Yes

ctO2([a-v]/a)

Yes

Hct

Yes

AG

Yes

AG(K)

Yes

cCa2+(7.4)

Yes

 

 

Auto QC

Yes

 

 

Sample type

Whole blood (arterial, venous), Serum, Plasma, Dialysate

Sampling method

Automatic aspiration.
No bio hazard risk because operators do not need to hold sample container during sampling.

 

 

Sample (uL)

 

Normal mode

195

Capillary mode

100

Micro mode

40

 

 

Analysis time (Sec)

About 85
NOTE: Cycle time is 180 – 250 sec.
System changes Flush time depending on concentration of Glc/Lac in blood sample

Display

Color display, touch panel

Operation for measurement

One touch

Entered parameters

Patient ID, Patient temperature, Sample type, FiO2, Height, Weight, Sex, Agge

Number of reagent

2 kinds (CAL Cartridge and FLUSH SOLUTION)

Shelf-life of reagent

9 months for both CAL Cartridge and FLUSH SOLUTION


SOLUTION

Calibration

Liquid calibration

Calibration frequency

1 point: Every 1 hour
2 point: Every 4 hour

Cleaning for protein

Automatic

Reagent detection

Barcode

Printer

Built-in, thermal printer

Interface

TCP/IP

barcode reader

Built-in

Power

85-264V, 50/60 Hz

Power consumption

240 VA

H x W x D (mm)

635 x 400 x 575

Weight (kg)

28

Battery

Built-in

Electrode Maintenance

Free

Other Maintenance

・Replace roller pump unit: Every 6month
・Replace sample port: Every 3month

Self diagnostic

Yes

Other function

・Enhanced data graphic (asid base, data transition) 
・GLAN Sever is available. (Remote control and data management system) 

Strong point


・Super Glc/Lac electrodes with on 3 months board shelf-life.
・Hands-free measurement with ideal sample container angle (Container at sampling module is designed to be laid at ideal angle to reduce sample contamination with air)